An ultrasensitive electrochemiluminescence (ECL) immunosensor for the detection of carcinoembryonic antigen (CEA) in serum and saliva was developed. Ru(bpy)(3)(2+)-graphene-Nafion composite was first coated on the glassy carbon electrode surface in this ECL immunosensor. The sandwich-type immunoreactions between the first antibody and the second antibody bridged the donors Ru(bpy)(3)(2+) and acceptors QDs, which led to the occurrence of ECL quenching of Ru(bpy)(3)(2+) by QDs. Based on the efficient ECL quenching, ultrasensitive CEA detection was realized. Under optimal conditions, the ECL signal depended linearly on the logarithm of the CEA concentration within a range from 0.005 to 0.5 pg/mL, and the detection limit of CEA was 0.002 pg/mL. Moreover, the proposed ECL immunosensor provides a stable, specific, and highly sensitive immunoassay protocol for CEA detection at very low levels, which might hold a promise for clinical application. (C) 2012 Elsevier B.V. All rights reserved.