4.7 Article

Potentiometric urea biosensor based on urease immobilized by an electrosynthesized poly(o-phenylenediamine) film with buffering capability

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 157, 期 1, 页码 211-215

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2011.03.051

关键词

Urea biosensor; Potentiometry; Electrochemical polymerization; Urease; Poly-o-phenylenediamine; Buffering capability

向作者/读者索取更多资源

A simple and novel potentiometric biosensor for urea detection was prepared by employing an electrosynthesized polymer with buffering capability. It was obtained by deposition of a weighed amount of urease (Ur) at a glassy carbon (GC) electrode followed by immobilization by an electrosynthesized poly-o-phenylenediamine (PPD) film. An unconventional upside-down (UD) geometry was employed for the electrochemical cell. The response of GC/Ur/PPD sensor is linear with urea concentration in the range 10 mu M to 1 mM (15 mV/mM, R-2 = 0.9999) due to buffering capability of PPD film, which represents a novel role of electrosynthesized polymers in their application to biosensors. At higher concentrations, the more common Nernstian response (28 mV/decade, R-2 = 0.9987) is observed. The sensor exhibits a sufficient sensitivity for practical determinations, rapid response and long term stability. (C) 2011 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据