4.7 Article

Amperometric glucose biosensor based on integration of glucose oxidase with platinum nanoparticles/ordered mesoporous carbon nanocomposite

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 153, 期 1, 页码 158-163

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2010.10.023

关键词

Ordered mesoporous carbon; Platinum nanoparticles; Biosensor; Glucose oxidase

资金

  1. Training Fund of NENU'S Scientific Innovation Project [NENU-STC08011]
  2. Science Foundation of Jilin Province [20090532]

向作者/读者索取更多资源

This article reports a new amperometric glucose biosensor based on ordered mesoporous carbon (OMC) supported platinum nanoparticles (Pt/OMC) modified electrode. The Pt/OMC nanocomposite modified electrode exhibited excellent electrocatalytic activities towards the reduction and oxidation of H2O2 as well. This feature allowed us to use it as bioplatform on which glucose oxidase (GOD) was immobilized by entrapment in electropolymerized pyrrole film for the construction of the glucose biosensor. The biosensor showed good analytical performances in terms of low detection (0.05 mM), high sensitivity (0.38 mu A/mM) and wide linear range (0.05-3.70 mM). In addition, the effects of pH value, applied potential, electroactive interference and the stability of the biosensor were discussed. The applicability to blood analysis was also evaluated. (C) 2010 Elsevier B.V. All rights reserved.