期刊
SENSORS AND ACTUATORS B-CHEMICAL
卷 141, 期 2, 页码 617-624出版社
ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2009.05.023
关键词
Cross-flow filtration; Centrifugation; Dean vortices; Cell-free layer; Recirculation flows
资金
- CEA
- Technologies for Health Programs (FAST project)
Our goal is to analyze and compare different continuous microfluidic principles dedicated to plasma extraction from hardly diluted human blood for lab-on-chip applications. First, the strengths and weaknesses of various emerging passive microfluidic methods (microfiltration- and centrifugation-based methods) were analyzed. Various devices were designed, microfabricated and tested with beads or blood. Filtration may be efficient, but with a high sample dilution, low flow rate and optimized geometry. Due to fast cell clogging, this remains a short-term solution. Separation effects resulting from centrifugal acceleration in curved channel flows are hindered by Dean vortices and anyhow are not pronounced with blood. An innovative device is then proposed and investigated experimentally. This is based on the lateral migration of red cells and the resulting cell-free layer, which is used to supply geometric singularities (an ear-cavity or a corner-edge) and locally enhance the clear plasma region. A maximum extraction of 10.7% is obtained for 1/20 diluted blood, injected at 100 mu L/min in the corner-edge design. (C) 2009 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据