期刊
SENSORS AND ACTUATORS B-CHEMICAL
卷 142, 期 1, 页码 210-215出版社
ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2009.08.029
关键词
Cu(II); QCM; Molecularly imprinted polymer (MIP); Determination
资金
- Henan Polytechnic University, China [B2009-78]
A novel quartz crystal microbalance (QCM) sensor with a high selectivity and sensitivity has been developed for the determination of Cu(II) ions, based on the modification of Cu(II) ion-imprinted polymer (Cu(II)-IIP) film onto a quartz crystal. The performance of the developed MIP-QCM sensor was evaluated and the results indicated that a sensitive MIP-QCM sensor could be fabricated. The obtained MIP-QCM sensor presents high-selectivity monitoring of Cu(II) ions, better reproducibility, shorter response time (6 min), wider linear range (0.001-50 mu M) and lower detection limit (8 x 10(-4) mu M). The practical analysis of the MIP-QCM sensor confirms the feasibility of Cu(II) determination in wastewater. (C) 2009 Elsevier B.V. All rights reserved.