An immunoelectrochemical sensor for salivary cortisol measurement

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 133, 期 2, 页码 533-537

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2008.03.018

关键词

salivary cortisol; immunoelectrochemical sensor; cyclic voltammetry; p-nitrophenyle phosphate

资金

  1. Graduate School in University of South Florida

向作者/读者索取更多资源

An immunoelectrochemical sensor based on alkaline phosphatase (AP) enzyme for determination of salivary cortisol concentration is reported. Microfabricated AU electrodes encased in a microfluidic chamber were functionalized to immobilize the cortisol capture antibodies. The reaction product p-nitrophenol (pNP) generated by reacting the AP enzyme attached to the cortisol antigen via detector antibodies with the p-nitrophenyl phosphate (pNPP) solution. pNP was detected as an oxidative peak between 0.9 and 1.1 V (vs. Ag pseudo-reference electrode) using cyclic voltammetry (CV) at room temperature. The magnitude of the peak varies linearly with the cortisol concentration, and was used to quantify the concentration of cortisol in real saliva samples. This immunoelectrochemical detection method accurately measured cortisol in the collected saliva samples achieved to a concentration of 0.76 nmol/L with an incubation time of 10 min. We demonstrate successfully the approach for establishing diurnal cortisol concentration behavior for clinical purposes with numerous advantages: a much higher throughput capability, significantly lower amounts of the sample, sub-pmol/L range sensitivity, higher resolution at low mass ranges, and easy to use. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据