4.7 Article

An optical biosensor for the rapid determination of glucose in human serum

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 129, 期 2, 页码 866-873

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2007.09.079

关键词

glucose optical biosensor; glucose oxidase; kinetic measurement; ORMOSIL-PVA; serum

向作者/读者索取更多资源

An optical glucose biosensor based on the immobilization of glucose oxidase in a sensing film was developed. The sensing film consisted of an organically modified silicate film embedded with tri(4,7-diphenyl-1,10-phenanthroline) ruthenium(H) perchlorate and a polyvinyl alcohol sol-gel matrix with immobilized glucose oxidase. A kinetic curve simulation method was employed for glucose measurements, as a result of which the measurement time was less than 1 min for each sample analysis. The detection limit (S/N = 3) of the glucose optical sensor was 3.6 x 10(-6) mol L-1 with linear ranges from 0.00 to 0.50 mmol L-1 (Y= 358.25X-2.93, R-2 = 0.9954) and 0.50 to 3.00 mmol L-1 (Y= 430.12X-46.43, R-2 = 0.9972). In addition, the appearance of the sensing film, and effects of the amounts of immobilized enzyme, pH, temperature, ionic strength, and co-existing substances (including heavy metal ions and organic compounds) were investigated. The glucose values estimated by this optical biosensor correlated well with those determined using the conventional method for human serum samples. (c) 2007 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据