4.7 Article

Ion chromatography-potentiometric detection of inorganic anions and cations using polypyrrole and overoxidized polypyrrole electrode

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 133, 期 1, 页码 5-14

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2008.01.004

关键词

polypyrrole; overoxidized polypyrrole; potentiometric detection; ion chromatography; water analysis

向作者/读者索取更多资源

Polypyrrole (PPy) and overoxidized polypyrrole (OPPy) films were used as a potentiometric detector in ion chromatography, in order to determine some anions and cations in water samples. Electrodeposition of PPy film was performed by constant potential at +0.80V (vs. Ag/AgCl) in an aqueous solution containing 0.1 M pyrrole and 0.2 M sodium nitrate. The overoxidation of PPy film was carried out in 0.1 M NaOH solution by cycling the potential between +0.80 and +1.20V at 50mV s(-1). The PPy film exhibited good sensitivity toward, especially, fluoride and chloride ions. The lowest detection limit (3 sigma) observed was 10(-8) M for fluoride with PPy detector. This is the first study using OPPy electrode for the potentiometric detection of both monovalent and divalent cations in literature. The correlation coefficients obtained from the calibration curves were 0.947-0.991 for anions and 0.998-0.999 for cations. The response of ions at the PPy and OPPy electrodes maintained over 90-94% of the original value for 45 and 30 days, respectively. It can be concluded from the observed data that these electrodes can be used as an alternative detector materials to analyse anions and cations in the water samples by ion chromatography. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据