期刊
SENSORS AND ACTUATORS A-PHYSICAL
卷 143, 期 1, 页码 90-96出版社
ELSEVIER SCIENCE SA
DOI: 10.1016/j.sna.2007.07.027
关键词
flexible neural probes; in vivo recording; tissue reaction
We present the fabrication, characterization, use in cortical recording and histological results of a flexible implantable neural probe. The microfabricated device is implemented in polyimide and platinum providing mechanical flexibility, high quality electrical characteristics and excellent biocompatibility. Two layers of platinum electrodes are used which greatly reduces the size of the neural probes thereby limiting the insertion damage. The quality of the individual recording sites was characterized using impedance spectroscopy. In recording experiments, acute in vivo measurements were performed in the mouse cortex. Local field potential, single- and multi-neuron activities were simultaneously recorded. We demonstrate using immumohistochemistry techniques a study of the inflammation at the neural probe implantation site. We show that the major advantage of using polymer probes over silicon probes is the reduced damage due to insertion and probe-brain compliance mismatch. (C) 2007 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐