4.6 Article

Temperature Stability of the Sky Quality Meter

期刊

SENSORS
卷 13, 期 9, 页码 12166-12174

出版社

MDPI AG
DOI: 10.3390/s130912166

关键词

light pollution; sky quality meter; light at night; temperature; stability; skyglow; night sky brightness

资金

  1. project Verlust der Nacht
  2. Federal Ministry of Education and Research, Germany [BMBF-033L038A]
  3. MILIEU (FU Berlin)

向作者/读者索取更多资源

The stability of radiance measurements taken by the Sky Quality Meter (SQM) was tested under rapidly changing temperature conditions during exposure to a stable light field in the laboratory. The reported radiance was found to be negatively correlated with temperature, but remained within 7% of the initial reported radiance over a temperature range of -15 degrees C to 35 degrees C, and during temperature changes of -33 degrees C/h and +70 degrees C/h. This is smaller than the manufacturer's quoted unit-to-unit systematic uncertainty of 10%, indicating that the temperature compensation of the SQM is adequate under expected outdoor operating conditions.