期刊
SENSORS
卷 13, 期 9, 页码 12166-12174出版社
MDPI AG
DOI: 10.3390/s130912166
关键词
light pollution; sky quality meter; light at night; temperature; stability; skyglow; night sky brightness
资金
- project Verlust der Nacht
- Federal Ministry of Education and Research, Germany [BMBF-033L038A]
- MILIEU (FU Berlin)
The stability of radiance measurements taken by the Sky Quality Meter (SQM) was tested under rapidly changing temperature conditions during exposure to a stable light field in the laboratory. The reported radiance was found to be negatively correlated with temperature, but remained within 7% of the initial reported radiance over a temperature range of -15 degrees C to 35 degrees C, and during temperature changes of -33 degrees C/h and +70 degrees C/h. This is smaller than the manufacturer's quoted unit-to-unit systematic uncertainty of 10%, indicating that the temperature compensation of the SQM is adequate under expected outdoor operating conditions.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐