4.6 Review

Rapid and Highly Sensitive Detection of Lead Ions in Drinking Water Based on a Strip Immunosensor

期刊

SENSORS
卷 13, 期 4, 页码 4214-4224

出版社

MDPI
DOI: 10.3390/s130404214

关键词

amplified competitive lateral flow assay; lead ions; immunosensor; gold nanoparticles

资金

  1. National Natural Science Foundation of China [21071066, 91027038, 21101079, 21175034]
  2. Key Programs from MOST [2012BAC01B07, 2012BAD29B05, 2012AA06A303, 2012BAD29B04, 2011BAK10B07, 2011BAK10B05, 2011BAK10B01, 2010AA06Z302, 2010DFB3047, 2011ZX08012-001, 2012BAK17B10, 2012BAK08B01, 2012YQ090194]
  3. Jiangsu Province
  4. MOF
  5. MOE [NCET-12-0879, BE2011626, 201210036, 201310135, 311002]

向作者/读者索取更多资源

In this study, we have first developed a rapid and sensitive strip immunosensor based on two heterogeneously-sized gold nanoparticles (Au NPs) probes for the detection of trace lead ions in drinking water. The sensitivity was 4-fold higher than that of the conventional LFA under the optimized conditions. The visual limit of detection (LOD) of the amplified method for qualitative detection lead ions was 2 ng/mL and the LOD for semi-quantitative detection could go down to 0.19 ng/mL using a scanning reader. The method suffered from no interference from other metal ions and could be used to detect trace lead ions in drinking water without sample enrichment. The recovery of the test samples ranged from 96% to 103%. As the detection method could be accomplished within 15 min, this method could be used as a potential tool for preliminary monitoring of lead contamination in drinking water.