Comparison of the Characteristics of Small Commercial NDIR CO2 Sensor Models and Development of a Portable CO2 Measurement Device

期刊

SENSORS
卷 12, 期 3, 页码 3641-3655

出版社

MDPI
DOI: 10.3390/s120303641

关键词

multipoint observation; CO2 sensor; calibration; portable CO2 measurement device

资金

  1. Ministry of Education, Science, Sports, and Culture [21310026, 20120005]
  2. Japan Society for the Promotion of the Science
  3. Grants-in-Aid for Scientific Research [20120005, 21310026] Funding Source: KAKEN

向作者/读者索取更多资源

Many sensors have to be used simultaneously for multipoint carbon dioxide (CO2) observation. All the sensors should be calibrated in advance, but this is a time-consuming process. To seek a simplified calibration method, we used four commercial CO2 sensor models and characterized their output tendencies against ambient temperature and length of use, in addition to offset characteristics. We used four samples of standard gas with different CO2 concentrations (0, 407, 1,110, and 1,810 ppm). The outputs of K30 and AN100 models showed linear relationships with temperature and length of use. Calibration coefficients for sensor models were determined using the data from three individual sensors of the same model to minimize the relative RMS error. When the correction was applied to the sensors, the accuracy of measurements improved significantly in the case of the K30 and AN100 units. In particular, in the case of K30 the relative RMS error decreased from 24% to 4%. Hence, we have chosen K30 for developing a portable CO2 measurement device (10 x 10 x 15 cm, 900 g). Data of CO2 concentration, measurement time and location, temperature, humidity, and atmospheric pressure can be recorded onto a Secure Digital (SD) memory card. The CO2 concentration in a high-school lecture room was monitored with this device. The CO2 data, when corrected for simultaneously measured temperature, water vapor partial pressure, and atmospheric pressure, showed a good agreement with the data measured by a highly accurate CO2 analyzer, LI-6262. This indicates that acceptable accuracy can be realized using the calibration method developed in this study.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据