期刊
SENSORS
卷 11, 期 8, 页码 7656-7664出版社
MDPI
DOI: 10.3390/s110807656
关键词
DNA sequence detection; quartz crystal microbalance; electronic oscillator; sensitivity; frequency noise; resolution
资金
- ANR PRECODD (HAB SEACHIP)
- European Union
This work deals with the design of a high sensitivity DNA sequence detector using a 50 MHz quartz crystal microbalance (QCM) electronic oscillator circuit. The oscillator circuitry is based on Miller topology, which is able to work in damping media. Calibration and experimental study of frequency noise are carried out, finding that the designed sensor has a resolution of 7.1 ng/cm(2) in dynamic conditions (with circulation of liquid). Then the oscillator is proved as DNA biosensor. Results show that the system is able to detect the presence of complementary target DNAs in a solution with high selectivity and sensitivity. DNA target concentrations higher of 50 ng/mL can be detected.
作者
我是这篇论文的作者