4.6 Article

Non-invasive glucose measurement by use of metabolic heat conformation method

期刊

SENSORS
卷 8, 期 5, 页码 3335-3344

出版社

MDPI
DOI: 10.3390/s8053335

关键词

sensor; heat dissipation; blood flow rate; degree of blood oxygen saturation; evaporation

向作者/读者索取更多资源

A non-invasive glucose measurement system based on the method of metabolic heat conformation (MHC) is presented in this paper. This system consists of three temperature sensors, two humidity sensors, an infrared sensor and an optical measurement device. The glucose level can be deduced from the quantity of heat dissipation, blood flow rate of local tissue and degree of blood oxygen saturation. The methodology of the data process and the measurement error are also analyzed. The system is applied in a primary clinical test. Compared with the results of a commercial automated chemistry analyzer, the correlation coefficient of the collected data from the system is 0.856. Result shows that the correlation coefficient improves when the factor of heat dissipated by evaporation of the skin is added in. A non-invasive method of measuring the blood flow rate of local tissue by heat transmission between skin and contacted conductor is also introduced. Theoretical derivation and numerical simulation are completed as well. The so-called normalized difference mean (NDM) is chosen to express the quantity of the blood flow rate. The correlation coefficient between the blood flow rates by this method and the results of a Doppler blood flow meter is equal to 0.914.