4.6 Article

Freestanding Cu nanowire arrays on Ti/Cr/Si substrate as tough nonenzymatic glucose sensors

期刊

RSC ADVANCES
卷 5, 期 101, 页码 82998-83003

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c5ra10058e

关键词

-

资金

  1. National Science Foundation of China (NSFC) [21203044, 31470489]
  2. Fundamental Research Funds for the Central Universities [HIT. IBRSEM. A. 201407]

向作者/读者索取更多资源

Exploring a tough, reusable and reproducible nonenzymatic sensor for glucose detection is required. Cu nanowire array electrodes on a Ti/Cr/Si substrate are prepared from an anodic aluminum oxide template film covered Ti/Cr/Si substrate by an electrodeposition method. The scanning electron microscope results indicate that Cu nanowires with a diameter of 50 nm and a length of hundreds of nm freestanding on the Ti/Cr/Si substrate are obtained. Electrochemical measurements indicate the freestanding nanowire array electrodes showing sensitivity of 1067 mu A mM(-1) cm(-2) and detection limit of 1.87 mu M when using the arrays as glucose sensors. The most attractive characteristics of the electrodes are the outstanding reusability and reproducibility, realizing a 1.77% relative standard deviation (RSD) of one electrode in five time tests and a 3.33% RSD of five electrodes for one sample. The application of the electrodes on determining the glucose concentration in human serum samples indicates the array electrode is a promising candidate as a practical sensor.