4.6 Article

Ultrasensitive electrochemical assay of hydrogen peroxide and glucose based on PtNi alloy decorated MWCNTs

期刊

RSC ADVANCES
卷 5, 期 124, 页码 102877-102884

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c5ra17410d

关键词

-

资金

  1. National Natural Science Foundation of China [21205030]
  2. Science and Technology Department of Hubei province [2014CFB548]
  3. State Key Laboratory of Advanced Technology for Materials Synthesis and Processing (Wuhan University of Technology) [2015-KF-13]
  4. Natural Science Fund for Creative Research Groups of Hubei Province of China [2014CFA015]

向作者/读者索取更多资源

A novel electrochemical assay based on PtNi/MWCNT nanocomposites was designed for ultrasensitive detection of hydrogen peroxide and glucose. The PtNi/MWCNT nanocomposites have been synthesized by a chemical reduction method and were investigated by X-ray diffraction, transmission electron microscopy, and electrochemical testing. The results demonstrated that the PtNi/MWCNT nanocomposites exhibited excellent electrocatalytic activity toward hydrogen peroxide and glucose. The electrocatalytic reduction of hydrogen peroxide showed a wide linear range from 0.2 mu M to 24.6 mM, with a high sensitivity of 2123.1 mu A mM(-1) cm(-2) and a detection limit of 60 nM (S/N = 3). At the same time, amperometric sensing of glucose was realized in a linear range from 0.1 mu M to 9.0 mM, with an excellent sensitivity of 85 910.0 mu A mM(-1) cm(-2). The present hybrid system also exhibited a good performance in terms of long-time stability, reproducibility and anti-interference ability.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据