期刊
RNA
卷 14, 期 3, 页码 584-592出版社
COLD SPRING HARBOR LAB PRESS, PUBLICATIONS DEPT
DOI: 10.1261/rna.761708
关键词
nicking endonuclease; dual-labeled fluorescent DNA probe; sequence-specific quantification; RNA quantification; conformation-interchangeable oligo-DNA; real-time quantification
We have developed a novel, single-step, isothermal, signal-amplified, and sequence-specific RNA quantification method (L-assay). The L-assay consists of nicking endonuclease, a dual-labeled fluorescent DNA probe (DL-probe), and conformation-interchangeable oligo-DNA (L-DNA). This signal-amplified assay can quantify target RNA concentration in a sequence-specific manner with a coefficient of variation (Cv) of 5% and a lower limit of detection of 0.1 nM. Moreover, this assay allows quantification of target RNA even in the presence of a several thousandfold excess by weight of cellular RNA. In addition, this assay can be used to measure the changes in RNA concentration in real-time and to quantify short RNAs (<30 nucleotides). The L-assay requires only incubation under isothermal conditions, is inexpensive, and is expected to be useful for basic research requiring high-accuracy, easy-to-use RNA quantification, and real-time quantification.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐