4.5 Article Proceedings Paper

A new neutron time-of-flight detector for fuel-areal-density measurements on OMEGA

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 85, 期 11, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.4886428

关键词

-

资金

  1. Department of Energy National Nuclear Security Administration [DE-NA0001944]
  2. University of Rochester
  3. New York State Energy Research and Development Authority
  4. DOE

向作者/读者索取更多资源

A new neutron time-of-flight (nTOF) detector for fuel-areal-density measurements in cryogenic DT implosions was installed on the OMEGA Laser System. The nTOF detector has a cylindrical thin-wall, stainless-steel, 8-in.-diam, 4-in.-thick cavity filled with an oxygenated liquid xylene scintillator. Four gated photomultiplier tubes (PMTs) with different gains are used to measure primary DT and D-2 neutrons, down-scattered neutrons in nT and nD kinematic edge regions, and to study tertiary neutrons in the same detector. The nTOF detector is located 13.4 m from target chamber center in a well-collimated line of sight. The design details of the nTOF detector, PMT optimization, and test results on OMEGA will be presented. (C) 2014 AIP Publishing LLC.