4.5 Article

Femtosecond broadband fluorescence upconversion spectroscopy: Improved setup and photometric correction

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 82, 期 6, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.3597674

关键词

-

资金

  1. Chinese Scholarship Council
  2. Deutsche Forschungsgemeinschaft [ER 154/9-1]

向作者/读者索取更多资源

A setup for fluorescence upconversion spectroscopy (FLUPS) is described which has 80 fs temporal response (fwhm) for emission in the spectral range 425-750 nm. Broadband phase matching is achieved with tilted gate pulses at 1340 nm. Background from harmonics of the gate pulse is removed and sensitivity increased compared to previous designs. Photometric calibration of the upconversion process is performed with a set of fluorescent dyes. For Coumarin 153 in methanol the peak position, bandwidth, and asymmetry depending on delay time are reported. (C) 2011 American Institute of Physics. [doi:10.1063/1.3597674]