4.5 Article

Instrument response standard in time-resolved fluorescence

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 80, 期 3, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.3095677

关键词

avalanche photodiodes; dyes; fluorescence; photon counting; standards; time resolved spectra; time resolved spectroscopy

资金

  1. Texas ETF (CCFT)
  2. THECB
  3. NSF [DBI-0649889]
  4. Polish Ministry of Science and Higher Education

向作者/读者索取更多资源

The fluorescence of LDS 798 dye in aqueous solution has a very short lifetime of 24ps, independent of excitation wavelength. The time response of common photon counting detectors depends on the wavelength of the registered photon. In lifetime measurements, the instrument response function (IRF) is usually approximated by the temporal profile of the scattered excitation light. Because lambda(Exc) is typically much shorter than lambda(Em), a systematic error may be present in these measurements. We demonstrate that the fluorescence decay of LDS 798 is a better approximation of IRF, in particular, for avalanche photodiodes used in the near infrared spectral region.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据