4.4 Article

Application of selected ion flow tube mass spectrometry to real-time atmospheric monitoring

期刊

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 24, 期 12, 页码 1763-1769

出版社

WILEY
DOI: 10.1002/rcm.4574

关键词

-

向作者/读者索取更多资源

Data are presented for real-time atmospheric monitoring of volatile organic chemicals (VOCs) in air using selected ion flow tube mass spectrometry (SIFT-MS) technology. These measurements were made by one of the new generation of SIFT-MS instruments. Results are shown for five VOCs that were continually monitored from a stationary sampling point over a 4-day period: ethene, ethanol, 1,3-butadiene, benzene and toluene. All analytes except ethene in the study have at least two simultaneous and independent measures of concentration. These results demonstrate the great advances in SIFT-MS that have been made in recent years. 1,3-Butadiene is measured at a concentration of 9 pptv with a precision of 44%. For a 1-s integration time, a detection limit of 50 pptv is achieved. Instrument sensitivities are reported for all five analytes. Copyright (C) 2010 John Wiley & Sons, Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据