On the calibration of continuous, high-precision δ18O and δ2H measurements using an off-axis integrated cavity output spectrometer

期刊

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 23, 期 4, 页码 530-536

出版社

WILEY
DOI: 10.1002/rcm.3905

关键词

-

资金

  1. Startup Grant from Princeton University
  2. The Office of Science (BER), US Department of Energy [DE-FG02-07ER64371]
  3. Directorate For Geosciences
  4. Division Of Earth Sciences [847368] Funding Source: National Science Foundation

向作者/读者索取更多资源

The O-18 and H-2 of water vapor serve as powerful tracers of hydrological processes. The typical method for determining water vapor delta O-18 and delta H-2 involves cryogenic trapping and isotope ratio mass spectrometry. Even with recent technical advances, these methods cannot resolve vapor composition at high temporal resolutions. In recent years, a few groups have developed continuous laser absorption spectroscopy (LAS) approaches for measuring delta O-18 and delta H-2 which achieve accuracy levels similar to those of lab-based mass spectrometry methods. Unfortunately, most LAS systems need cryogenic cooling and constant calibration to a reference gas, and have substantial power requirements, making them unsuitable for long-term field deployment at remote field sites. A new method called Off-Axis Integrated Cavity Output Spectroscopy (OA-ICOS) has been developed which requires extremely low-energy consumption and neither reference gas nor cryogenic cooling. In this report, we develop a relatively simple pumping system coupled to a dew point generator to calibrate an ICOS-based instrument (Los Gatos Research Water Vapor Isotope Analyzer (WVIA) DLT-100) under various pressures using liquid water with known isotopic signatures. Results show that the WVIA can be successfully calibrated using this customized system for different pressure settings, which ensure that this instrument can be combined with other gas-sampling systems. The precisions of this instrument and the associated calibration method can reach similar to 0.08 parts per thousand. for delta O-18 and similar to 0.4 parts per thousand for delta H-2. Compared with conventional mass spectrometry and other LAS-based methods, the OA-ICOS technique provides a promising alternative tool for continuous water vapor isotopic measurements in field deployments. Copyright (C) 2009 John Wiley & Sons, Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据