Improved imaging resolution in desorption electrospray ionization mass spectrometry

期刊

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 22, 期 17, 页码 2639-2644

出版社

WILEY-BLACKWELL
DOI: 10.1002/rcm.3662

关键词

-

向作者/读者索取更多资源

The imaging resolution of desorption electrospray ionization mass spectrometry (DESI-MS) was investigated using printed patterns on paper and thin-layer chromatography (TLC) plate surfaces. Resolution approaching 40 mu m was achieved with a typical DESI-MS setup, which is approximately 5 times better than the best resolution reported previously. This improvement was accomplished with careful control of operational parameters (particularly spray tip-to-surface distance, solvent flow rate, and spacing of lane scans). In addition, an appropriately strong analyte/surface interaction and uniform surface texture on the size scale no larger than the desired imaging resolution were required to achieve this resolution. Overall, conditions providing the smallest possible effective desorption/ionization area in the DESI impact plume region and minimizing the analyte redistribution on the surface during analysis led to improved DESI-MS imaging resolution. Published in 2008 by John Wiley & Sons, Ltd.