4.4 Article

Role of the support material on laser desorption/ionization mass spectra

期刊

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 22, 期 7, 页码 925-929

出版社

JOHN WILEY & SONS LTD
DOI: 10.1002/rcm.3425

关键词

-

向作者/读者索取更多资源

We report the results of experimental studies on the effects of sample supports in laser desorption/ionization mass spectrometry (LDI-MS). LDI time-of-flight (TOF) mass spectra obtained for C-60 and insulin samples deposited onto standard stainless steel substrate and/or onto some non-metallic materials (glass, scotch tape, floppy disc foil, Teflon foil, photocopy film), all recorded under identical, typical experimental conditions, have been compared with regard to their intensity and quality. The LDI investigations show that compared with stainless steel, glass and floppy disc foil sample supports boost (2-3.5 times) ion yields for C-60(+) and C-60(-) ions, respectively. The stainless steel and scotch tape sample supports are the best for the mass resolution of positive ions and the formation of (C-60)(n)(-) (n <= 4) cluster ions, respectively. In the case of detection of insulin by matrix-assisted laser desorption/ionization (MALDI) we did not observe significant differences in sensitivity for the support materials tested. A mechanism of ion formation in the desorption plume is suggested. Copyright (C) 2008 John Wiley & Sons, Ltd.