4.7 Article

Evaluation of Heavily Calcified Vessels with Coronary CT Angiography: Comparison of Iterative and Filtered Back Projection Image Reconstruction

期刊

RADIOLOGY
卷 260, 期 2, 页码 390-399

出版社

RADIOLOGICAL SOC NORTH AMERICA
DOI: 10.1148/radiol.11103574

关键词

-

资金

  1. Department of Veterans Affairs

向作者/读者索取更多资源

Purpose: To prospectively compare traditional filtered back projection (FBP) and iterative image reconstruction for the evaluation of heavily calcified arteries with coronary computed tomography (CT) angiography. Materials and Methods: The study had institutional review board approval and was HIPAA compliant. Written informed consent was obtained from all patients. Fifty-five consecutive patients (35 men, 20 women; mean age, 58 years +/- 12 [standard deviation]) with Agatston scores of at least 400 underwent coronary CT angiography and cardiac catheterization. Image data were reconstructed with both FBP and iterative reconstruction techniques with corresponding cardiac algorithms. Image noise and subjective image quality were compared. To objectively assess the effect of FBP and iterative reconstruction on blooming artifacts, volumes of circumscribed calcifications were measured with dedicated volume analysis software. FBP and iterative reconstruction series were independently evaluated for coronary artery stenosis greater than 50%, and their diagnostic accuracy was compared, with cardiac catheterization as the reference standard. Statistical analyses included paired t tests, Kruskal-Wallis analysis of variance, and a modified McNemar test. Results: Image noise measured significantly lower (P = .011-.035) with iterative reconstruction instead of FBP. Image quality was rated significantly higher (P = .031 and .042) with iterative reconstruction series than with FBP. Calcification volumes measured significantly lower (P = .019 and .026) with iterative reconstruction (44.3 mm(3) +/- 64.7 and 46.2 mm(3) +/- 68.8) than with FBP (54.5 mm(3) +/- 69.5 and 56.3 mm(3) +/- 72.5). Iterative reconstruction significantly improved some measures of per-segment diagnostic accuracy of coronary CT angiography for the detection of significant stenosis compared with FBP (accuracy: 95.9% vs 91.8%, P = .0001; specificity: 95.8% vs 91.2%, P = .0001; positive predictive value: 76.9% vs 61.1%, P = .0001). Conclusion: Iterative reconstruction reduces image noise and blooming artifacts from calcifications, leading to improved diagnostic accuracy of coronary CT angiography in patients with heavily calcified coronary arteries. (C) RSNA, 2011

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据