Prostate Tumor Volume Measurement with Combined T2-weighted Imaging and Diffusion-weighted MR: Correlation with Pathologic Tumor Volume

期刊

RADIOLOGY
卷 252, 期 2, 页码 449-457

出版社

RADIOLOGICAL SOC NORTH AMERICA
DOI: 10.1148/radiol.2523081423

关键词

-

资金

  1. National Institutes of Health [R01 CA76423]

向作者/读者索取更多资源

Purpose: To retrospectively determine the accuracy of diffusion-weighted (DW) magnetic resonance (MR) imaging for identifying cancer in the prostate peripheral zone (PZ) and to assess the accuracy of tumor volume measurements made with T2-weighted imaging and combined T2-weighted and DW MR imaging by using surgical pathologic examination as the reference standard. Materials and Methods: The institutional review board issued a waiver of informed consent for this HIPAA-compliant study. Forty-two patients underwent endorectal MR at 1.5 T before undergoing radical prostatectomy for prostate cancer and had at least one PZ tumor larger than 0.1 cm(3) at surgical pathologic examination. On T2-weighted images, an experienced radiologist outlined suspected PZ tumors. Two apparent diffusion coefficient (ADC) cutoff values were identified by using the Youden index and published literature. Image cluster analysis was performed on voxels within the suspected tumor regions. Associations between volume measurements from imaging and from pathologic examination were assessed by using concordance correlation coefficients (CCCs). The sensitivity and specificity of ADCs for identifying malignant PZ voxels were calculated. Results: In identifying malignant voxels, respective ADC cutoff values of 0.0014 and 0.0016 mm(2)/sec yielded sensitivity of 82% and 95% and specificity of 85% and 65%, respectively. Sixty PZ cancer lesions larger than 0.1 cm(3) were found at pathologic examination; 43 were detected by the radiologist. CCCs between imaging and pathologic tumor volume measurements were 0.36 for T2-weighted imaging, and 0.46 and 0.60 for combined T2-weighted and DW MR imaging with ADC cutoffs of 0.0014 and 0.0016 mm2/sec, respectively; the CCC of combined T2-weighted and DW MR imaging (ADC cutoff, 0.0016 mm2/sec) was significantly higher (P = .006) than that of T2-weighted imaging alone. Conclusion: Adding DW MR to T2-weighted imaging can significantly improve the accuracy of prostate PZ tumor volume measurement. (C) RSNA, 2009

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据