Delayed Gadolinium-enhanced MR Imaging of Articular Cartilage: Three-dimensional T1 Mapping with Variable Flip Angles and B1 Correction

期刊

RADIOLOGY
卷 252, 期 3, 页码 865-873

出版社

RADIOLOGICAL SOC NORTH AMERICA
DOI: 10.1148/radiol.2531081115

关键词

-

向作者/读者索取更多资源

Purpose: To develop and verify the accuracy of a rapid imaging protocol for delayed gadolinium-enhanced magnetic resonance (MR) imaging of cartilage that was based on three-dimensional (3D) spoiled gradient-recalled acquisition in the steady state (SPGR) sequences with variable flip angles (FAs) (VFAs) and where a correction method for B-1 field inhomogeneities was applied. Materials and Methods: The institutional research ethics board approved this study. Written informed consent was obtained from all subjects. A B1 field inhomogeneity correction method was applied to a 3D SPGR pulse sequence with VFAs (repetition time msec/echo time msec, 7.1/3.3; FAs, 2 degrees, 5 degrees, 10 degrees, and 20 degrees) and was used to perform delayed gadolinium-enhanced MR imaging of cartilage 3D T1 measurements at 1.5 T. The 3D T1 measurements were validated with the reference standard ( the results of T1 mapping by using a single-section two-dimensional [2D] inversion-recovery [IR] fast spin-echo [SE] pulse sequence in vitro and in vivo) in six healthy volunteers. Results: T1 values calculated from 3D T1 maps were not significantly different from reference T1 values in vitro (P = .195) and in vivo (P = .52) when a B1 field inhomogeneity correction method was applied. In vivo T1 mapping of the articular surface of the whole femoropatellar joint, including data acquisition, was performed in approximately 8 minutes of acquisition time at a spatial resolution of 0.55 x 0.55 x 3.00 mm. Conclusion: Rapid T1 mapping by using 3D SPGR acquisitions with a VFA approach and a correction for B1 field inhomogeneities can be used for delayed gadolinium-enhanced MR imaging of cartilage. T1 measurements performed in vitro and in vivo by using this approach are highly accurate when compared with those performed by using standard 2D IR fast SE T1 mapping as a reference.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据