4.7 Article

Pulmonary nodule volume: Effects of reconstruction parameters on automated measurements - A phantom study

期刊

RADIOLOGY
卷 247, 期 2, 页码 400-408

出版社

RADIOLOGICAL SOC NORTH AMERICA
DOI: 10.1148/radiol.2472070868

关键词

-

资金

  1. NCI NIH HHS [K24 CA120494] Funding Source: Medline

向作者/读者索取更多资源

Purpose: To prospectively evaluate in a phantom the effects of reconstruction kernel, field of view (FOV), and section thickness on automated measurements of pulmonary nodule volume. Materials and Methods: Spherical and lobulated pulmonary nodules 3-15 mm in diameter were placed in a commercially available lung phantom and scanned by using a 16-section computed tomographic (CT) scanner. Nodule volume (V) was determined by using the diameters of 27 spherical nodules and the mass and density values of 29 lobulated nodules measured by using the formulas V = (4/3) pi r3 ( spherical nodules) and V = 1000 center dot (M/D) (lobulated nodules) as reference standards, where r is nodule radius; M, nodule mass; and D, wax density. Experiments were performed to evaluate seven reconstruction kernels and the independent effects of FOV and section thickness. Automated nodule volume measurements were performed by using computer-assisted volume measurement software. General linear regression models were used to examine the independent effects of each parameter, with percentage overestimation of volume as the dependent variable of interest. Results: There was no substantial difference in the accuracy of volume estimations across the seven reconstruction kernels. The bone reconstruction kernel was deemed optimal on the basis of the results of a series of statistical analyses and other qualitative findings. Overall, volume accuracy was significantly associated (P < .0001) with larger reference standard-measured nodule diameter. There was substantial overestimation of the volumes of the 3-5- mm nodules measured by using the volume measurement software. Decreasing the FOV facilitated no significant improvement in the precision of lobulated nodule volume measurements. The accuracy of volume estimations-particularly those for small nodules-was significantly (P < .0001) affected by section thickness. Conclusion: Substantial, highly variable overestimation of volume occurs with decreasing nodule diameter. A section thickness that enables the acquisition of at least three measurements along the z-axis should be used to measure the volumes of larger pulmonary nodules.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据