4.2 Article

National survey of patient doses from whole-body FDG PET-CT examinations in France in 2011

期刊

RADIATION PROTECTION DOSIMETRY
卷 152, 期 4, 页码 334-338

出版社

OXFORD UNIV PRESS
DOI: 10.1093/rpd/ncs066

关键词

-

资金

  1. French Society of Nuclear Medicine

向作者/读者索取更多资源

A national survey of patient doses from owhole-body PET-CT' examinations was conducted within all French nuclear medicine departments in 2011. Data related to injected [F-18]-fluorodeoxyglucose (FDG) activity and to computerised tomography (CT) parameters were received from 56 positron emission tomography (PET)-CT units (answer rate: approximate to 60 ). The average specific injected FDG activity was equal to 4.3 MBq kg(1), in agreement with European recommendations. The new otime-of-flight' technology enabled to decreasing the specific activity down to 3.5 MBq kg(1). The results have shown that current diagnostic reference levels (DRLs) for the diagnostic trunk CT are too high for CT combined with PET, only performed in France for attenuation correction and localisation, and not for diagnostic purpose. Despite wide variations between PET-CT units (4-fold factor in CTDIvol), DRLs equal to 8 mGy (CTDIvol) and 750 mGy cm (dose-length product) could be proposed for whole-body PET-CT. The average effective dose related to whole-body PET-CT examination in France has been assessed to approximate to 14 mSv.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据