Use of active personal dosemeters in interventional radiology and Tests in hospitals - ORAMED project

期刊

RADIATION MEASUREMENTS
卷 46, 期 11, 页码 1258-1261

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.radmeas.2011.08.002

关键词

Active personal dosemeters; Interventional radiology; Interventional cardiology; Hospitals

向作者/读者索取更多资源

Although active personal dosemeters (APDs) are not used quite often in hospital environments, the possibility to assess the dose and/or dose rate in real time is particularly interesting in interventional radiology and cardiology (IR/IC) since operators can receive relatively high doses while standing close to the primary radiation field. A study concerning the optimization of the use of APDs in IR/IC was performed in the framework of the ORAMED project, a Collaborative Project (2008-2011) supported by the European Commission within its 7th Framework Program. This paper reports on tests performed with APDs on phantoms using an X-ray facility in a hospital environment and APDs worn by interventionalists during routine practice in different European hospitals. The behaviour of the APDs is more satisfactory in hospitals than in laboratories with respect to the influence of the tube peak high voltage and pulse width, because the APDs are tested in scattered fields with dose equivalent rates generally lower than 1 Sv.h(-1). (C) 2011 Elsevier Ltd. All rights reserved.