INTERNAL STANDARD VERSUS EXTERNAL STANDARD CALIBRATION: AN UNCERTAINTY CASE STUDY OF A LIQUID CHROMATOGRAPHY ANALYSIS

期刊

QUIMICA NOVA
卷 33, 期 4, 页码 984-987

出版社

SOC BRASILEIRA QUIMICA
DOI: 10.1590/S0100-40422010000400041

关键词

internal and external standard calibration; measurement uncertainty; bias correction factor

向作者/读者索取更多资源

Traditionally, in the cigarettes industry, the determination of ammonium ion in the mainstream smoke is performed by ion chromatography. This work studies this determination and compares the results of this technique with the use of external and internal standard calibration. A reference cigarette sample presented measurement uncertainty of 2.0 mu g/cigarette and 1.5 mu g/cigarette, with external and internal standard, respectively. It is observed that the greatest source of uncertainty is the bias correction factor and that it is even more significant when using external standard, confirming thus the importance of internal standardization for this correction.