Over the last 30 years, IUPAC published several documents with the goal of achieving standardized nomenclature and methodology for potentiometric ion-selective electrodes (ISEs). The ISE vocabulary was formulated, measurement protocols were suggested, and the selectivity coefficients were compiled. However, in liaht of new discoveries and experimental possibilities in the field of ISEs, some of the IUPAC recommendations have become outdated. The goal of this technical report is to direct attention to ISE practices and the striking need for updated or refined IUPAC recommendations which are consistent with the state of the art of using macro- and microfabricated planar microelectrodes. Some of these ISE practices have never been addressed by IUPAC but have gained importance with the technological and theoretical developments of recent years. In spite of its recognized importance, a generally acceptable revision of the current IUPAC recommendations is far beyond the scope of this work.