PROSTHETICS AND ORTHOTICS INTERNATIONAL
卷 39, 期 3, 页码 255-259出版社
SAGE PUBLICATIONS LTD
DOI: 10.1177/0309364613520319
关键词
Feedback; upper-leg prostheses; error-based
资金
- Dutch Technology Foundation STW of the Netherlands Organization for Scientific Research (NWO)
- Ministry of Economic Affairs, Agriculture and Innovation [08003]
Background and aim: Several studies have shown that feedback in upper-leg prostheses is possible, but slow or difficult to interpret. In this study, electrotactile and auditive error-based feedback, only giving feedback when an undesired event occurs, were tested for its use in upper-leg prosthesis when provided during a perturbation. Technique: A total of nine healthy subjects walked on a prosthetic simulator which was disturbed at the end of the swing phase. They received either no feedback, electrotactile feedback, or auditive feedback at the time of the perturbation. Discussion: The reaction time of the subjects only improved by 40 ms when using auditory feedback, compared to the no-feedback condition. No changes in reaction time were found in the electrotactile feedback condition. Considering perturbation detection was not taken into account in this study, this improvement is not enough for practical applications in upper-leg prosthesis. Clinical relevance Many transfemoral amputees are insecure about their prosthesis, are afraid of falling, or actually fall. Providing feedback specifically during a perturbation may prevent them from falling, or at least give them a chance to react.