- National Science Foundation
A transportable prototype Faraday rotation spectroscopic system based on a tunable external cavity quantum cascade laser has been developed for ultrasensitive detection of nitric oxide ( NO). A broadly tunable laser source allows targeting the optimum Q(3/2)(3/2) molecular transition at 1875.81 cm(-1) of the NO fundamental band. For an active optical path of 44 cm and 1-s lock-in time constant minimum NO detection limits (1 sigma) of 4.3 parts per billion by volume (ppbv) and 0.38 ppbv are obtained by using a thermoelectrically cooled mercury-cadmium-telluride photodetector and liquid nitrogen-cooled indium-antimonide photodetector, respectively. Laboratory performance evaluation and results of continuous, unattended monitoring of atmospheric NO concentration levels are reported.