4.2 Article

The stability of a hip fracture determines the fatigue of an intramedullary nail

出版社

PROFESSIONAL ENGINEERING PUBLISHING LTD
DOI: 10.1243/09544119JEIM664

关键词

Gamma nail; failure; finite element method; fatigue; hip fracture

向作者/读者索取更多资源

The purpose of this study was to address the question of how the stability of a proximal hip fracture determines the fatigue and failure mechanism of an intramedullary implant. To answer this question, mechanical experiments and finite element simulations with two different loading scenarios were conducted. The two load scenarios differed in the mechanical support of the fracture by an artificial bone sleeve, representing the femoral head and neck. The experiments confirmed that an intramedullary nail fails at a lower load in an unstable fracture situation in the proximal femur than in a stable fracture. The nails with an unstable support failed at a load 28 per cent lower than the nails with a stable support by the femoral neck. Hence, the mechanical support of a fracture is crucial to the fatigue failure of an implant. The simulation showed why the fatigue fracture of the nail starts at the aperture of the lag screw. It is the location of the highest von Mises stress, which is the failure criterion for ductile materials.