Infrared (IR) modalities represent the only currently viable mass fever screening approaches for outbreaks of infectious disease pandemics such as Ebola virus disease and severe acute respiratory syndrome. Non-contact IR thermometers (NCITs) and IR thermographs (IRTs) have been used for fever screening in public areas such as airports. While NCITs remain a more popular choice than IRTs, there has been increasing evidences in the literature that IRTs can provide great accuracy in estimating body temperature if qualified systems are used and appropriate procedures are consistently applied. In this study, we addressed the issue of IRT qualification by implementing and evaluating a battery of test methods for objective, quantitative assessment of IRT performance based on a recent international standard (IEC 80601-2-59). We tested two commercial IRTs to evaluate their stability and drift, image uniformity, minimum resolvable temperature difference, and radiometric temperature laboratory accuracy. Based on these tests, we illustrated how experimental and data processing procedures could affect results, and suggested methods for clarifying and optimizing test methods. Overall, the insights into thermograph standardization and acquisition methods provided by this study may improve the utility of IR thermography and aid in comparing IRT performance, thus improving the potential for producing high quality disease pandemic countermeasures.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据