4.6 Article

A Multimodal, SU-8-Platinum - Polyimide Microelectrode Array for Chronic In Vivo Neurophysiology

期刊

PLOS ONE
卷 10, 期 12, 页码 -

出版社

PUBLIC LIBRARY SCIENCE
DOI: 10.1371/journal.pone.0145307

关键词

-

资金

  1. Bolyai Janos Grant of the HAS
  2. KTIA [NAP 13-2-2015-0004]
  3. OTKA [K81354]
  4. ANR-TET Multisca
  5. EU FP7 NeuroSeeker grants [600925]
  6. [KTIA_13_NAP-A-IV/1-2-3-6]
  7. [TAMOP-4.2.1.B-11/2/KMR-2011-0002]

向作者/读者索取更多资源

Utilization of polymers as insulator and bulk materials of microelectrode arrays (MEAs) makes the realization of flexible, biocompatible sensors possible, which are suitable for various neurophysiological experiments such as in vivo detection of local field potential changes on the surface of the neocortex or unit activities within the brain tissue. In this paper the microfabrication of a novel, all-flexible, polymer-based MEA is presented. The device consists of a three dimensional sensor configuration with an implantable depth electrode array and brain surface electrodes, allowing the recording of electrocorticographic (ECoG) signals with laminar ones, simultaneously. In vivo recordings were performed in anesthetized rat brain to test the functionality of the device under both acute and chronic conditions. The ECoG electrodes recorded slow-wave thalamocortical oscillations, while the implanted component provided high quality depth recordings. The implants remained viable for detecting action potentials of individual neurons for at least 15 weeks.