Quantification of Shear-Induced Platelet Activation: High Shear Stresses for Short Exposure Time

期刊

ARTIFICIAL ORGANS
卷 39, 期 7, 页码 576-583

出版社

WILEY-BLACKWELL
DOI: 10.1111/aor.12438

关键词

Blood-shearing device; Shear-induced platelet activation; Shear stress; Exposure time; Power law model

资金

  1. National Institutes of Health [R01 HL 088100]

向作者/读者索取更多资源

Thrombosis and thromboembolism are the life-threatening clinical complications for patients supported or treated with prosthetic cardiovascular devices. The high mechanical shear stress within these devices is believed to be the major contributing factor to cause platelet activation (PA) and function alteration, leading to thrombotic events. There have been limited quantitative data on how the high mechanical shear stress causes platelet activation. In this study, shear-induced PA in the ranges of well-defined shear stress and exposure time relevant to cardiovascular devices was quantitatively characterized for human blood using two novel flow-through Couette-type blood shearing devices. Four markers of platelet activation-surface P-selectin (CD62p), platelet-derived microparticles (PMPs), platelet-monocyte aggregation (PMA), and soluble P-selectin-were measured by flow cytometry and enzyme-linked immunosorbent assay (ELISA), respectively. The results indicated that PA induced by high shear stresses with short exposure time could be reliably detected with surface P-selectin, and, to a lesser extent, PMPs rather than soluble P-selectin. It was also verified that PMA can be a highly sensitive indirect marker of platelet activation. The quantitative relationship between percentage of activated platelets indicated by surface P-selectin expression and shear stress/exposure time follows well the power law functional form. The coefficients of the power law models of PA based on surface P-selectin expression were derived.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据