PHYSIOLOGICAL MEASUREMENT
卷 36, 期 1, 页码 N23-N34出版社
IOP PUBLISHING LTD
DOI: 10.1088/0967-3334/36/1/N23
关键词
data acquisition; anti-aliasing filter; low-pass filter; biomedical signal processing; digital filter simulation
The availability of microcomputer-based portable devices facilitates the high-volume multichannel biosignal acquisition and the analysis of their instantaneous oscillations and inter-signal temporal correlations. These new, non-invasively obtained parameters can have considerable prognostic or diagnostic roles. The present study investigates the inherent signal delay of the obligatory anti-aliasing filters. One cycle of each of the 8 electrocardiogram (ECG) and 4 photoplethysmogram signals from healthy volunteers or artificially synthesised series were passed through 100-80-60-40-20 Hz 2-4-6-8th order Bessel and Butterworth filters digitally synthesized by bilinear transformation, that resulted in a negligible error in signal delay compared to the mathematical model of the impulse- and step responses of the filters. The investigated filters have as diverse a signal delay as 2-46 ms depending on the filter parameters and the signal slew rate, which is difficult to predict in biological systems and thus difficult to compensate for. Its magnitude can be comparable to the examined phase shifts, deteriorating the accuracy of the measurement. As a conclusion, identical or very similar anti-aliasing filters with lower orders and higher corner frequencies, oversampling, and digital low pass filtering are recommended for biosignal acquisition intended for inter-signal phase shift analysis.