4.5 Article

Telemetry pill versus rectal and esophageal temperature during extreme rates of exercise-induced core temperature change

期刊

PHYSIOLOGICAL MEASUREMENT
卷 33, 期 6, 页码 915-924

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0967-3334/33/6/915

关键词

core temperature; gastrointestinal temperature; exertional hyperthermia; telemetry; thermometry

向作者/读者索取更多资源

Core temperature measurement with an ingestible telemetry pill has been scarcely investigated during extreme rates of temperature change, induced by short high-intensity exercise in the heat. Therefore, nine participants performed a protocol of rest, (sub)maximal cycling and recovery at 30 degrees C. The pill temperature (T-pill) was compared with the rectal temperature (T-re) and esophageal temperature (T-es). T-pill corresponded well to T-re during the entire trial, but deviated considerably from T-es during the exercise and recovery periods. During maximal exercise, the average Delta T-pill-T-re and Delta T-pill-T-es were 0.13 +/- 0.26 and - 0.57 +/- 0.53 degrees C, respectively. The response time from the start of exercise, the rate of change during exercise and the peak temperature were similar for T-pill and T-re. T-es responded 5 min earlier, increased more than twice as fast and its peak value was 0.42 +/- 0.46 degrees C higher than T-pill. In conclusion, also during considerable temperature changes at a very high rate, T-pill is still a representative of T-re. The extent of the deviation in the pattern and peak values between T-pill and T-es (up to > 1 degrees C) strengthens the assumption that T-pill is unsuited to evaluate central blood temperature when body temperatures change rapidly.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据