4.5 Article Proceedings Paper

Frequency-difference MIT imaging of cerebral haemorrhage with a hemispherical coil array: numerical modelling

期刊

PHYSIOLOGICAL MEASUREMENT
卷 31, 期 8, 页码 S111-S125

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0967-3334/31/8/S09

关键词

magnetic induction tomography; haemorrhagic cerebral stroke

资金

  1. EPSRC [EP/E009697/1] Funding Source: UKRI
  2. Engineering and Physical Sciences Research Council [EP/E009697/1] Funding Source: researchfish

向作者/读者索取更多资源

The feasibility of detecting a cerebral haemorrhage with a hemispherical MIT coil array consisting of 56 exciter/sensor coils of 10 mm radius and operating at 1 and 10 MHz was investigated. A finite difference method combined with an anatomically realistic head model comprising 12 tissue types was used to simulate the strokes. Frequency-difference images were reconstructed from the modelled data with different levels of the added phase noise and two types of a priori boundary errors: a displacement of the head and a size scaling error. The results revealed that a noise level of 3 m degrees (standard deviation) was adequate for obtaining good visualization of a peripheral stroke (volume approximate to 49 ml). The simulations further showed that the displacement error had to be within 3-4 mm and the scaling error within 3-4% so as not to cause unacceptably large artefacts on the images.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据