Simultaneous imaging using Si-PM-based PET and MRI for development of an integrated PET/MRI system

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 57, 期 2, 页码 N1-N13

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/57/2/N1

关键词

-

资金

  1. Ministry of Education, Science, Sports and Culture, Japan
  2. National Institute of Biomedical Innovation, Osaka, Japan
  3. Grants-in-Aid for Scientific Research [23659598, 21390351, 21591561] Funding Source: KAKEN

向作者/读者索取更多资源

The silicon photomultiplier (Si-PM) is a promising photo-detector for PET for use in magnetic resonance imaging (MRI) systems because it has high gain and is insensitive to static magnetic fields. Recently we developed a Si-PM-based depth-of-interaction PET system for small animals and performed simultaneous measurements by combining the Si-PM-based PET and the 0.15 T permanent MRI to test the interferences between the Si-PM-based PET and an MRI. When the Si-PM was inside the MRI and installed around the radio frequency (RF) coil of the MRI, significant noise from the RF sequence of the MRI was observed in the analog signals of the PET detectors. However, we did not observe any artifacts in the PET images; fluctuation increased in the count rate of the Si-PM-based PET system. On the MRI side, there was significant degradation of the signal-to-noise ratio (S/N) in the MRI images compared with those without PET. By applying noise reduction procedures, the degradation of the S/N was reduced. With this condition, simultaneous measurements of a rat brain using a Si-PM-based PET and an MRI were made with some degradation in the MRI images. We conclude that simultaneous measurements are possible using Si-PM-based PET and MRI.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据