4.6 Article

Dual-energy attenuation coefficient decomposition with differential filtration and application to a microCT scanner

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 55, 期 4, 页码 1141-1155

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/55/4/016

关键词

-

资金

  1. National Institutes of Health [NIH-NCI 2U24 CA092865]

向作者/读者索取更多资源

Dual-energy x-ray computed tomography (DECT) has the capability to decompose attenuation coefficients using two basis functions and has proved its potential in reducing beam-hardening artifacts from reconstructed images. The method typically involves two successive scans with different x-ray tube voltage settings. This work proposes an approach to dual-energy imaging through x-ray beam filtration that requires only one scan and a single tube voltage setting. It has been implemented in a preclinical microCT tomograph with minor modifications. Retrofitting of the microCT scanner involved the addition of an automated filter wheel and modifications to the acquisition and reconstruction software. Results show that beam-hardening artifacts are reduced to noise level. Acquisition of a mu-Compton image is well suited for attenuation-correction of PET images while dynamic energy selection (4D viewing) offers flexibility in image viewing by adjusting contrast and noise levels to suit the task at hand. All dual-energy and single energy reference scans were acquired at the same soft tissue dose level of 50 mGy.