The influence of CT image noise on proton range calculation in radiotherapy planning

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 55, 期 6, 页码 N141-N149

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/55/6/N01

关键词

-

向作者/读者索取更多资源

The purpose of this note is to evaluate the relationship between the stochastic errors in CT numbers and the standard deviation of the computed proton beam range in radiotherapy planning. The stochastic voxel-to-voxel variation in CT numbers called 'noise,' may be due to signal registration, processing and numerical image reconstruction technique. Noise in CT images may cause a deviation in the computed proton range from the physical proton range, even assuming that the error due to CT number-stopping power calibration is removed. To obtain the probability density function (PDF) of the computed proton range, we have used the continuing slowing down approximation (CSDA) and the uncorrelated white Gaussian noise along the proton path. The model of white noise was accepted because for the slice-based fan-beam CT scanner; the power-spectrum properties apply only to the axial (x, y) domain and the noise is uncorrelated in the z domain. However, the possible influence of the noise power spectrum on the standard deviation of the range should be investigated in the future. A random number generator was utilized for noise simulation and this procedure was iteratively repeated to obtain convergence of range PDF, which approached a Gaussian distribution. We showed that the standard deviation of the range, sigma, increases linearly with the initial proton energy, computational grid size and standard deviation of the voxel values. The 95% confidence interval width of the range PDF, which is defined as 4 sigma, may reach 0.6 cm for the initial proton energy of 200 MeV, computational grid 0.25 cm and 5% standard deviation of CT voxel values. Our results show that the range uncertainty due to random errors in CT numbers may be significant and comparable to the uncertainties due to calibration of CT numbers.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据