4.6 Article Proceedings Paper

A feasibility study of treatment verification using EPID cine images for hypofractionated lung radiotherapy

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 54, 期 18, 页码 S1-S8

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/54/18/S01

关键词

-

向作者/读者索取更多资源

We propose a novel approach for potential online treatment verification using cine EPID (electronic portal imaging device) images for hypofractionated lung radiotherapy based on a machine learning algorithm. Hypofractionated radiotherapy requires high precision. It is essential to effectively monitor the target to ensure that the tumor is within the beam aperture. We modeled the treatment verification problem as a two-class classification problem and applied an artificial neural network (ANN) to classify the cine EPID images acquired during the treatment into corresponding classes-with the tumor inside or outside of the beam aperture. Training samples were generated for the ANN using digitally reconstructed radiographs (DRRs) with artificially added shifts in the tumor location-to simulate cine EPID images with different tumor locations. Principal component analysis (PCA) was used to reduce the dimensionality of the training samples and cine EPID images acquired during the treatment. The proposed treatment verification algorithm was tested on five hypofractionated lung patients in a retrospective fashion. On average, our proposed algorithm achieved a 98.0% classification accuracy, a 97.6% recall rate and a 99.7% precision rate.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据