4.6 Article

A tandem calibration method for kerma-area product meters

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 53, 期 18, 页码 4941-4958

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/53/18/006

关键词

-

向作者/读者索取更多资源

For measurements of the kerma-area product (KAP) in diagnostic x-ray imaging, a method for calibrating field KAP meters with a reference KAP meter is presented. In this tandem calibration method, the field KAP chamber is positioned similarly as in measurements with patients. The reference KAP chamber is placed at a specified distance and used in the x-ray beam simultaneously with the field KAP chamber. The tandem method provides a feasible and practical way for calibrating field KAP meters of any type in their clinical position. Accurate measurements of the irradiation geometry are not required, but comprehensive calibration for the reference KAP meter is needed.