4.5 Article

METHOD FOR ESTIMATING THE ACOUSTIC PRESSURE IN TISSUES USING LOW-AMPLITUDE MEASUREMENTS IN WATER

期刊

ULTRASOUND IN MEDICINE AND BIOLOGY
卷 41, 期 11, 页码 3001-3012

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2015.07.005

关键词

Liver attenuation; Porcine liver; Non-linear distortion; Drug delivery; In situ pressure; Exposimetry; Derating

资金

  1. European Regional Development Fund
  2. Republic of Cyprus through Research Promotion Foundation [DIDAKTOR/0311/78]

向作者/读者索取更多资源

The aim of this study was to evaluate a simple, reliable and reproducible method for accuracy in estimating the acoustic pressure delivered in tissue exposed to ultrasound. Such a method would be useful for therapeutic applications of ultrasound with microbubbles, for example, sonoporation. The method is based on (i) low-amplitude water measurements that are easily made and do not suffer from non-linear propagation effects, and (ii) the attenuation coefficient of the tissue of interest. The range of validity of the extrapolation method for different attenuation and pressure values was evaluated with a non-linear propagation theoretical model. Depending on the specific tissue attenuation, the method produces good estimates of pressures in excess of 10 MPa. Ex vivo machine-perfused pig liver tissue was used to validate the method for source pressures up to 3.5 MPa. The method can be used to estimate the delivered pressure in vivo in diagnostic and therapeutic applications of ultrasound. (E-mail: maverk@uw.edu) (C) 2015 World Federation for Ultrasound in Medicine & Biology.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据