3.9 Review

Intricacies of Dose in Laser Phototherapy for Tissue Repair and Pain Relief

期刊

PHOTOMEDICINE AND LASER SURGERY
卷 27, 期 3, 页码 387-393

出版社

MARY ANN LIEBERT, INC
DOI: 10.1089/pho.2009.2503

关键词

-

类别

向作者/读者索取更多资源

Inaccurate measurement and incorrect reporting of dosages are major shortcomings of phototherapy articles. As many as 30% of published reports in the field either lack relevant information needed to determine a dosage or report dosages that are altogether inaccurate. The high prevalence of dosage-related mistakes in published reports suggests that dosage determination errors are common among clinicians and other end-users. This special article is designed to advance understanding of the relevant parameters used in phototherapy for tissue repair and pain relief, particularly among clinicians and others who may not be completely familiar with the technology. I define and discuss five key parameters that influence dosage, including 1) radiant power, 2) radiant energy, 3) power density, 4) energy density, and 5) wavelength, and use hypothetical cases to demonstrate how factors such as beam spot size, size of lesion, mode of treatment (contact, noncontact, or scanning), frequency of treatment, dose per treatment, and cumulative dose affect dosages and treatment outcomes. The potential effects of patient-related factors, such as etiology, pathology, tissue optical density, depth of target tissue, and skin pigmentation are discussed concurrently and strategies are suggested to improve dosage determination.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.9
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据