Performance benchmarks of an array-based hand-held photoacoustic probe adapted from a clinical ultrasound system for non-invasive sentinel lymph node imaging

出版社

ROYAL SOC
DOI: 10.1098/rsta.2010.0353

关键词

photoacoustic imaging; clinical ultrasound array system; sentinel lymph node biopsy; image-guided needle biopsy

资金

  1. National Institutes of Health [R01 EB000712, R01 EB008085, R01 CA134539, U54 CA136398, R01 EB010049, 5P60 DK02057933]

向作者/读者索取更多资源

Clinical translation of photoacoustic (PA) imaging can be facilitated by integration with commercial ultrasound (US) scanners to enable dual-modality imaging. An array-based US scanner was modified for hand-held PA imaging. The performance was benchmarked in terms of signal-to-noise ratio (SNR), axial spatial resolution and sensitivity. PA images of a tube, filled with methylene blue (MB; approx. 30mM) and placed at various depths in chicken tissue, were acquired. A 5 cm penetration depth was achieved with an 18.6 dB SNR using a laser fluence of 3mJcm(-2), only one-seventh of the safety limit (20mJcm(-2)). An axial resolution of approximately 400 mu m was maintained at all imaging depths. The PA sensitivity to MB placed 2.3 cm deep in chicken tissue was less than 100 mu M. Further, after intradermal injection of MB (approx. 30 mM), a rat sentinel lymph node was clearly identified in vivo, beneath a 3.8 cm thick layer of chicken breast. The accumulated concentration of MB in the node was estimated to be approximately 7mM. The noise-equivalent sensitivities (approx. 2 cm depth) were 17 and 85 mu M, ex vivo and in vivo, respectively. These results support the use of this PA system for non-invasive mapping and image-guided needle biopsy of sentinel nodes in breast cancer patients.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据