Multivariate calibration of NIR spectroscopic sensors for continuous glucose monitoring

期刊

TRAC-TRENDS IN ANALYTICAL CHEMISTRY
卷 67, 期 -, 页码 147-158

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.trac.2014.12.005

关键词

Blood-glucose measurement; Chemometrics; Continuous glucose monitoring; Diabetes; In-vivo use; Multivariate calibration; NIR spectroscopy; Pre-processing strategy; Sensor; Variable selection

资金

  1. I.W.T. Flanders [IWT 100021, SB-090053]

向作者/读者索取更多资源

Diabetes, a disorder in the control of blood-glucose levels, is one of the most serious metabolic diseases worldwide. Among the investigated technologies for continuous glucose monitoring (CGM), near-infrared spectroscopy (NIR) has received the most attention. There have been many attempts to develop NIR-based CGM systems with promising in-vitro results, but they lacked robustness for in-vivo use. We critically review the application of chemometrics for CGM and the research needed. Pre-processing and multivariate-calibration techniques, which allow exploiting expert knowledge on the potential interferences, are possible solutions. The combination and first overtone bands in the ranges 2050-2300 nm and 1500-1800 nm, respectively, are the most informative regions. We therefore recommended selecting the most informative variables and exploiting the available expert knowledge on known interferences in preprocessing or multivariate calibration to develop an NIR-based CGM sensor for in-vivo use. (C) 2014 Elsevier B.V. All rights reserved.